The NECSA_Brucker_D8_Advance Instrument

Brucker D8 Advance lab powder diffratometer

Identification

  • Site: NECSA
  • Author: us all in ZA !
  • Origin: Necsa, ZA
  • Date: 2024 Oct 11th

Description

This is a lab-scale diffractometer, Brucker D8 Advance type.

- X-ray Tubes Cu (1.54 Å), Co (1.79 Å), Cr (2.29 Å)
- Beam sizes 0.1 x 15 mm2; 1 mm collimator
- Distance btw souurce and 1st slit: 332mm
- sample-detector distance: 435 mm
- Detector LynxEye 1D 192 pixels 14.4 x 16 mm

Example: 

Input parameters

Parameters in boldface are required; the others are optional.
NameUnitDescriptionDefault
source_materialstranode type, among Cu.txt, Co.txt, Cr.txt"Cu.txt"
EminkeVminimum energy at the source1
EmaxkeVmaximum energy at the source9
Gobel_radiusmGobel curved mirror curvature radius0.5
Gobel_rotationdegGobel curved mirror rotation20
Gobel_coatingstrGobel curved mirror coating, e.g. Ir.dat"Ir.dat"
Sample_materialstrsample material, e.g. CIF/LAU/LAZ file"Si.laz"

Links

  • Source code for NECSA_Brucker_D8_Advance.instr.
  • http://nanoqam.ca/wiki/lib/exe/fetch.php?media=d8_advance_discover_user_manual_vol._1_doc-m88-exx153_v6.pdf